Categories
Categories
Brands
Brands
Defibtech Lifeline View Semi-Auto AED with LCD screen External Defibrillator, 4 year battery, 8 yr warranty
Defibtech Lifeline View Semi-Auto AED with LCD screen External Defibrillator, 4 year battery, 8 yr warranty
Product Description
Lifeline VIEW Tech Specs Please note "Special Price" the pads maybe on or under the two year expiry date
DDU-2300 Series Defibrillator
TYPE
Semi Automatic External Defibrillator
WAVEFORM
Biphasic Truncated Exponential
(Impedance compensated)
ENERGY
Adult: 150 Joules
Child / Infant: 50 Joules
(Nominal into 50 Ohm load)
CONTROLS
1. Lighted On/Off button
2. Lighted Shock button
CHARGE TIME *
Less than 4 seconds
(shock decision to shock)
DISPLAY
High-resolution color LCD
VIDEO PROMPTS
Full motion video
On-screen text prompts
CPR COACHING
Video and voice coaching
On-demand video help
VOICE PROMPTS
Extensive voice prompts guide user through operation of the unit
RESCUE PROTOCOL
AHA 2005
Supports protocol updates by the user
(password protected)
* Typical, with new battery at 25° C
WARRANTY
8 year Warranty
Self Tests
Automatic daily, weekly and monthly circuitry tests
BATTERY INSERTION
System integrity test on battery insertion
PAD PRESENCE
Pads pre-connected tested daily
USER-INITIATED
Unit and battery pack system test initiated by the user
STATUS INDICATION
Visual and audible indication of unit status
STATUS SCREEN
Unit self-test results
Pads and battery information
(status and expiration)
Patient Analysis System
PATIENT ANALYSIS
Automatically evaluates patient impedance for proper pad contact. Monitors signal quality and analyzes patient ECG for shockable/non-shockable rhythms.
SENSITIVITY/SPECIFICITY
Meets AAMI-DF-80 specifications and AHA recommendations
Defibrillation / Monitoring Pads
MODEL
Adult - DDP-2001
Child / Infant - DDP-2002
TYPE
Pre-connected, single-use, non-polarized, disposable, self-adhesive electrodes with cable and connector
SURFACE AREA
77 cm2 (nominal, each pad)
50 cm2 (nominal, each pad)
Environmental TEMPERATURE
Operating: 0 to 50°C
Standby: 0 to 50°C
RELATIVE HUMIDITY
Operating / Standby: 5 (non-condensing)
ALTITUDE
-150 to 4500m per MIL-STD-810F 500.4 Procedure ll
VIBRATION
Ground (MIL-STD-810F 514.5 Category 20)
Helicopter (RTCA/DO-160D, Section 8.8.2, Cat R. Zone 2, Curve G)
Jet Aircraft (RTCA/DO-160D Section 8. Cat H, Zone 2, Curves B & R)
SHOCK / DROP ABUSE TOLERANCE
MIL-STD-810F 516.5 Procedure IV 48 inches (1.2 meters), any edge, corner, or surface, in standby mode
CRUSH TEST
450 kg
SEALING / WATER RESISTANCE
IEC60529 class IP55; Protected against dust and waterjets (battery pack installed)
ESD
EN61000-4-2: (Open air up to 8kV or direct contact up to 6kV)
EMC (Emission)
CISPR 11 Group 1 Level B
and FCC Part 15
EMC (Immunity)
IEC 61000-4-3 and IEC 61000-4-8
Battery Pack
MODEL DBP-2300 (4 years)
POWER
12V, 2800 mAh
Lithium/Manganese Dioxide Disposable, recyclable, non-rechargeable
CAPACITY
125 shocks or 8 hours continuous operation
STANDBY LIFE *
4 years
LOW BATTERY INDICATORS
Visible
Audible
* Typical, with new battery at 25° C
Event Documentation
INTERNAL EVENT RECORD
Critical ECG segments and rescue event parameters are recorded (greater than 60 minutes) and can be downloaded to a removable data card
PC-BASED EVENT REVIEW
ECG with event tag display, and audio playback when available
REMOVABLE STORAGE
(optional) Up to 30 hours of ECG and event data storage (no audio option) or up to 3 hours of audio (audio option). ECG and event storage on a removable data card. Actual length of storage is dependent on card capacity.
USB PORT
Event download and maintenance operations
SIZE
18.5 x 24 x 5.8 cm
WEIGHT
Less than 1.4 kg with battery